The results obtained with spiked mice and sheep blood samples, using an external calibration in the range of 1-120 mg dL-1ethanol, gave recoveries of 93.2-104.4% (n = 12). The "Blood Alcohol Micro-pad" gave good precision with %RSD less then 1 (50 mg dL-1 ethanol, n = 1 and limit of quantification (10SD of intercept/slope) of 11.56 mg dL-1. The method was successfully validated against a headspace gas chromatography-mass spectrometric method. It has good potential for development as a simple and convenient blood alcohol se