https://www.selleckchem.com/pr....oducts/Temsirolimus.
The current approach to Linac beam dosimetry verification is typically performed utilizing a three-dimensional (3D) water tank system. The 3D beam scanning process is cumbersome, labor intensive, error-prone, and costly. This is especially challenging for the new Ethos system and MR Linacs with a ring gantry. This work proposes an alternative approach to verify 6FFF beam dosimetry for Ethos, ViewRay MRIdian® Linac, and other Linacs with 6FFF beam quality using two-dimensional (2D) ion chamber arrays. Percentage depth dose (PDD) and