6 week ago - Translate

https://www.selleckchem.com/pr....oducts/l-name-hcl.ht
All measurements are expressed in absolute values. Four-hundred-seventy-six TTs and 96 OFRs were analyzed. The intra-class correlation coefficient calculated for the calibrated thermal mass flowmeter was  0.99 and reflected the excellent reliability of our measurements. For TTs, the bias value was - 0.24 L/min (± 0.88), and the limits of agreement were - 1.97 to 1.48 L/min. For OFRs, the bias value was - 0.30 L/min (± 0.54), and the limits of agreement were - 1.36 to 0.77 L/min. As the flow increased, the accuracy of all analyzed